9 Maintenance Matters of Handheld Brinell Hardness Tester

9 Maintenance Matters of Handheld Brinell Hardness Tester

1. The tester should follow the operation procedure and be able to proofread the handheld Brinell hardness tester with standard blocks before and after the test.
2. Handheld Brinell hardness tester, which is not often used, should be tested several times in standard after start-up, and then tested again after stabilization.
3. The use of hardness blocks can only be carried out on the working surface, and the distance between test points in each test is not less than 3 mm.
4. In the hardness test, it is strictly forbidden to rotate the load-changing handwheel after applying the main test force.
5. When carrying the hardness tester, the measuring lever should be fixed first, and the weights and suspenders should be taken off. If the weights, suspenders and upper cover are taken off, or the fuses need to be replaced, the power plugs should be pulled out first.
6. The handheld Brinell hardness tester should be kept clean, and dust-proof cover should be put on after the test. Hardness block and ball pressing head should be coated with anti-rust oil to prevent rust after use.
7. The hardness tester needs periodic verification, at least once a year, which ensures the accuracy of the hardness tester.
8. Troubleshooting of common faults of hardness tester: When the hardness tester fails, we should contact the relevant units and repair it. Generally, common fixing can be solved by ourselves.
9. When the handheld Brinell hardness tester is not used for a long time, the machine needs to be covered with a dust-proof cover.

Regularly it needs to be injected with a small amount of oil into the contact surface between the lead screw and the handwheel. Before using the hardness tester, the top face of the screw and the end face of the worktable should be wiped clean.

If it is found that the hardness indication error is large:
1. The worktable can be removed to check whether the contact surface between the worktable and the screw is clean.
2. Check whether the protective sleeve of the lead screw lifts the worktable;
3. Check whether the pressure head is damaged.



Related News
  • The Use Function of UV Light Meter

    The Use Function of UV Light Meter

    April 8, 20201. Easy to use: Simply connect the new sensor, and the host engine will automatically configure it without entering new programs or calibration factors. The detector settings and algorithms are compre...view
  • Brief introduction of Portable Color Meter

    Brief introduction of Portable Color Meter

    May 14, 2019Product introduction:Portable color meter is suitable for large, medium and small waterworks, industrial and mining enterprises, domestic or industrial water colorimetric detection, in order to contro...view
  • Brief Introduction of Gloss Meter

    Brief Introduction of Gloss Meter

    May 14, 20191. Introduction of products:This instrument is a precise optical instrument. Please keep and use the instrument properly. It should be avoided to use and store the instrument in humid, strong electrom...view
  • Introduction of Handheld Brinell Hardness Tester

    Introduction of Handheld Brinell Hardness Tester

    June 11, 2019Performance characteristics of hand-held brinell hardness tester- Stable load, determined by pin shear force of 1580kg, independent of operator- High repeatability, comparable to desktop computers- Lo...view