UV Light Meter Test Ultraviolet Power UV Intensity Energy For UVA UVB UVC Support 7 Probe With Water Proof Sensor LS125
-
The ultraviolet band including UVA (320nm-380nm), UVB band (280nm-320nm) and UVC band (200nm-280nm).
-
UVA wave band is mainly used for light curing, UV light exposure lithography, lithography, integrated circuit industry etc..
-
UVC band, also known as the short wave ultraviolet sterilization. What ultraviolet germicidal lamp issued is short ultraviolet band UVC.
-
LS125 multi probe UV radiometer, using digital probe, plug design, instrument judge probe model intelligently. At the same time can support a variety of UV test probe, currently supports the following 7 kinds of UV Probes.
Paremeters
1. Instrument Weight:about 320g
2. Main Body Size:125mm × 72mm × 14.5mm (L*W*H)
3. Power supply: 2 AAA alkaline battery
4. Probe connection: push pull (Snap-in style)
Probe Parameters
1: P254 Probe
1) Response wavelength: 230nm-280nm; λp=254nm
2) Measuring range:0 - 200000 uW/cm2
3) Resolution: 0.1 μW/cm2
4) Accuracy: ±10 %
5) Test Hole Diameter: Φ10
6) Probe Size: Diameter 46mm×Thickness 16.1mm
2: P254-wp Probe
1) Response wavelength: 230nm-280nm; λp=254nm
2) Measuring range:0 - 200000 uW/cm2
3) Resolution: 0.1 μW/cm2
4) Accuracy: ±10 %
5) Test Hole Diameter: Φ10
6) Probe Size: Diameter 50mm×Thickness 20mm
7) Waterproof depth:1 meter
3: P297 Probe
1) Response wavelength: 280nm-320nm;
suitable for measure UVB light source that peak wavelength at 297nm,308nm,313nm etc.
2) Measuring range:0 - 200000 uW/cm2
3) Resolution: 0.1 μW/cm2
4) Accuracy: ±10 %
5) Test Hole Diameter: Φ10
6) Probe Size: Diameter 46mm×Thickness 16.1mm
4: P365L Probe
1) Response wavelength:260nm-400nm; λp=365nm
2) Measuring range:0-200000uW/cm2
3) Accuracy: ±10 %
4) Resolution: 0.1 uW/cm2
5) Test Hole Diameter:Φ10
6) Probe Size: Diameter 46mm×Thickness 12.2mm
5: P420L Probe
1) Response wavelength:340nm-420nm;λp=395nm
2) Measuring range:0-200000uW /cm2
3) Accuracy: ±10 %
4) Resolution: 0.1 uW /cm2
5) Test Hole Diameter:Φ9.2
6) Probe Size: Diameter 46mm×Thickness 14.6mm
6: E365 Probe
1) Response wavelength:320nm-400nm; λp=365nm
2) Power Measuring range:0 - 10000 mW/cm2
3) Energy Measuring range: 0 --- 999999mJ/cm2
4) Accuracy: ±10 %
5) Resolution: 0.1 mW/cm2
6) Test Hole Diameter:Φ9.2
7) Probe Size: Diameter 46mm×Thickness 14.6mm
7: E395 Probe
1) Response wavelength:340nm-420nm;λp=395nm
2) Power Measuring range:0-20000mW/cm2
3) Energy Measuring range: 0 --- 999999mJ/cm2
4) Accuracy: ±10 %
5) Resolution: 0.1 mW/cm2
6) Test Hole Diameter:Φ9.2